Digital precision quick-response thermometer with freely adjustable analog
output
GMH 3210
Battery / mains operation for plug-in probes, data hold function, freely
adjustable analog output 0 to1V
| Applications: Suitable for all quick-response,
high-precision temperature measurements. Analog output 0-1V for recorder
connection, min/max. memory.
Special features:
- freely adjustable analog output or serial interface, selectable by
menu setting
- Min/Max-value memory, hold function
- Correction of meas. value for surface measurement can be switched
on/off.
- Zero-point offset entry
- 5 different thermocouples selectable (Type K, J,N, S or T)
- Battery and d. c. operation
- Automatic power off, selectable within 1 to120min, or permanent
operation
- Low power consumption of ca. 0.3 mA (with included standard battery
ca. 1,000 hours of operation)
- Display in °C or °F (selectable)
- ready (for up to 5 instruments of the GMH3000-series can be
connected to a single serial PC interface)
|

|
(without accessories)
Specification:
| Thermocouples:
|
Type K, J, N, S, T (acc. to DIN EN
60584) |
| Measuring
ranges:
(extract)
|
Type K: (NiCr-Ni) |
-65,0..+300.0°C resp. -220...+1372°C
(-85,0 ... +572,0°F resp. -364
... +2500°F) |
|
Type S: (Pt10Rh-Pt) |
-50 ... +1768°C (-58 ...+3214°F) |
|
Type N: (NiCrSi-NiSi) |
-100,0 ... +380,0°C bzw. -200
... +1300°C
(-148,0 ... +482,0°F resp. -328
... +2372°F)
|
| Resolution: |
0.1°C or 1°C (0.1°F or 1°F)
|
|
Accuracy:
(extract)
|
+/- 1
Digit at nominal temperature
|
|
Type K:
|
-65.0 to 300.0°C : + 0.03%
of meas. val.. + 0.05% FS
-220 to 1372 °C: + 0.08%
of meas. val. + 0.1% FS
|
|
Type S:
|
-50 to 1768°C: + 0.1%
of meas. val. + 0.1% FS
|
|
Type N:
|
-100,0 ... +380,0 + 0.03%
of meas. val.. + 0.05% FS
-200,0 ... +1300,0 + 0.08%
of meas. val.. + 0.1% FS (T>-100°C)
-200,0 ... +1300,0 + 1°C
of meas. val.. + 0.1% FS (T<-100°C) |
| Temperature
drift: |
0.01%/K
|
| Point of
comparison: |
+ 0.3°C
|
| Analog
output:
|
0 to 1V, freely adjustable,
resolution 13 bit, accuracy 0.05% at nominal temperature |
| Nominal
temperature: |
25°C
|
| Working
temperature: |
-25 to +50°C
|
| Relative
humidity: |
0 to 95% (non-condensing)
|
| Storage
temperature: |
-25 to +70°C
|
| Probe
connection: |
jacks for the flat pin plug for the
simultaneous connection |
| Display: |
2 four digit LCDs for temperature,
min/max-values, hold function etc. as well as additional functional arrows
|
| Min/Max-value
memory: |
Memorizing of max. and min.
values
|
| Hold function: |
By pressing a button the
current values are memorized.
|
| Push
buttons: |
6 membrane keys for ON/OFF-switch,
feature selection, min-max-value memory, hold-function
|
| Power
supply: |
9V battery, Type IEC6F22 (included) as well as additional d. c. connector
(internal pin diam. 1.9 mm) for external 10.5-12V direct voltage supply. (suitable
power supply: GNG 10/3000) |
| Low battery
warning: |
! and "bAT"
|
| Power
consumption: |
ca. 0.3 mA
|
| Auto-Off-Function:: |
By activating the auto off function the device
is automatically switched off, if its interface and
keypad is not attended for a longer time (selectable 1 to120min) |
| Interface: |
serial interface (choice
between interface or analog output operation is done via menu), direct
connection to RS232 interface or USB3100 interface of a PC via electrically isolated interface
adapter GRS 3100 / GRS 3105.
|
| Dimensions: |
142x71x26 mm (HxWxD)
Impact-resistant ABS plastic housing, membrane keyboard, transparent panel.
Front side IP65, integrated pop-up clip
|
| Special
applications:
|
Compensation for surface
measurements:
A compensation value (to
compensate for loss when transferring heat from the meas. object to the
probe) can be set and switched on/off for surface measurements, if
required.
Zero-point offset entry:
By entering the offset
temperature the parameter can be moved parallely to the calibration graph.
|
Accessories:
| GNG 10/3000 |
net power supply |
| GB 9V |
spare battery |
| GKK 3000 |
case with punched lining |
| GKK 3100 |
case with foam lining |
| GKK 3500 |
large case with punched lining |
| ST-N1 |
device protection bag with 1 right angled hole |
| ST-N2 |
device protection bag with 2 right angled holes |
| GRS 3100 |
interface converter with 1 connection |
| GRS 3105
|
interface converter with 2 connections |
| USB 3100 |
interface converter to USB |
| EBS 9M |
software for transmission |
| probes
type K |
(up to 1150°C) |
| probes
type N |
(up to 1300°C) |
| probes
type S |
(up to 1600°C) |
Digital precision quick-response thermometer with freely adjustable analog output or serial interface (sensor not incl.)
Please order the following accessories separately:
Please order furthermore separately:
= data sheet as *.pdf
= instruction manual as *.pdf
deutsche Beschreibung
Price of the basic article, German V.A.T. incl
:
EUR
146.50
The net price of the basic article GMH 3210
is:
EUR
123.11
|