Pt100- High Precision Thermometer
Refernece meter for any calibration requirement Microprozessor-Precision
thermometer for Pt100 4-wire
|
GMH 3710
|
- suitable for all Pt100 4-wire probes with 4-pin miniature DIN-plug
- highest accuracy and resolution (0,01°C)
- freely adjustable analog output 0-1V or seriell interface
- Offset and solpe input
- Min-/Max-value memory, hold function
|
  |
GMH 3750
|
additional functions:
- two integrated logger functions
- optical and acoustic min-max-alarm
- user defined sensor curve (50 interpolation points)
- real-time clock with date and year
|
Application:
Reference measurings in liquids, soft media, air/gases
|
Measuring
range: |
-199,99 ... +199,99 °C
resp. -200,0 .... + 850,0 °C
-199,99 .....+199,99 °F resp. -328,0 .... 1562,0 °F
|
|
Resolution: |
0,01 °C resp. 0,1 °C
0,01 °F resp. 0,1 °F
|
|
Linearisation: |
digital stored characteristic
curve (DIN EN 6075)1 - GMH 3750
add. supports an userdefined curve.
|
|
Auto-Range: |
automatically or manually choose
of the measuring range
|
|
Accuracy: |
(at nominal temperature = 25°C)
< +0,03 °C (+0,06 °F) at meas. range -199,99..199,99°C
|
|
Temperature drift: |
< 0,002 °C/K
|
|
Probe: |
Pt100, 4-wire, according to DIN EN 60751
probe connection via 4-pin miniature plug
|
|
Nominal
temperature: |
25 °C
|
|
Working
temperature:
|
-25 ... +50 °C
|
|
Relative
Humidity: |
0 .... 95 % r.F. (non condensing)
|
|
Storage
temperature: |
-25 .... +70 °C
|
|
Display: |
two 41/2
digit LCDs (12,4mm resp. 7 mm high),
as well as additional arrows
|
|
Pushbuttons: |
6 membrane keys
|
|
output: |
3-pin jack connector 3,5 mm
diameter, choice between seriell interface or analog output
- seriell interface: direct connection to RS 232 resp. USB -interface
of a PC via electrically isolated interface adapter GRS 3100 or
GRS 3105 resp.
USB 3100 (p.r.t. accessories)
- analog output: 0-1V, freely
adjustable (resolution 13 bit,
accuracy 0.05% at nominal temperature)
|
|
power
supply: |
9V-battery, type IEC 6F22 (included)
as well as additional d. c. connector for external 10.5-12V direct
voltage supply (suitable power supply:
GNG 10/3000)
|
|
low
battery warning: |
automatically 'bAt'
|
|
power
consumption: |
approx. 1mA
|
|
Dimensions: |
142 x 71 x 26 mm (H x W x D)
|
|
Weight: |
approx. 155 g
|
|
min-max-value: |
memorizing of max. and min.
values
|
|
Hold function: |
By pressing a button the current
values will be "frozen"
|
|
automatic-Off-funktion: |
1 ... 120 min (can also be
deactivated)
|
|
Offset and slope input: |
offset- and scale correction can
be entered digitally
|
|
min-/max-alarm: |
the measuring value is constantly
monitored if they remain within the min./max. limits
- Alarm: 3 different alarm settings:
off:
alarm function not activated
on:
visual alarm via display, integrated buzzer and interface
no Sound: alarm via display and
interface
- regulating function:
with the help of the switching module GAM 3000 (optionally)
electric equipment can be switched on/off or alarm memoried
|
|
Logger
functions: |
- manually:
99 data sets (data recall via keyboard or interface)
- cycle:
16.200 data sets (data recall via interface)
-
adjustable cycle time:
1 sec to 1 h
Logger start and stop via the keyboard or interface. Comfortable read-out
and display software (GSOFT
3050) available as additional equipment. )
|
|
Real-time clock: |
clock with date and year |
Accessories:
 |
GMHKonfig |
| Software description: Comfortable software
to edit the userdefined sensor curve of the GMH 3750 (e.g. for
calibration laboratories...) By means of this software probes can be
adjusted to the instrument. As result an overall accuracy of <0,03°C can
be achieved depending of the measuring range. Similar resistance curves
(e.g. Ni100) can be loaded to. The sensor curve can be stored external
and reloaded to the device. Receivable upon request!
calibrated systems:

General:
Te overall error of a measuring consists of the sum of the instrument
error and the probe error. To minimise the overall error, we offer
calibrated and optimised systems below. Due to their excellent system
accuracy they are especially suitable for quality assurance according to
ISO9000ff, as reference instruments in manufacturing processes, laboratory,
service and maintenance, etc.
The system optimisation is done via a special characteristic curve which
is determined for each temperature probe separately and stored in the
instrument.
|
GMH 3750 SET2
| Scope of delivery: |
Meas. device GMH 3750, temperature probe GTF 401,
case GKK 3500 and certificate of calibration with three
calibration points (opt. meas. range:
-20..+250°C) |
| accuracy: |
better than 0,3°C |
| Calibration points: |
0°C/100°C/250°C |
|
GMH 3750 SET1
| Scope of delivery: |
Measuring device GMH 3750, temperature probe GTF 401,
plastic case GKK 3500 and certificate of calibration with 3
calibration points (opt. Meas. range: -20..+70°C) |
| Accuracy: |
better than 0,07°C |
| Calibration points: |
-20°C/0°C/70°C |
|
GMH 3750 SET3
| Scope of delivery: |
Meas. device GMH 3750, temperature probe GTF 601,
case GKK 3500 and certificate of calibration with three
calibration points (opt. meas. range:
-20..+500°C) |
| Accuracy: |
better than 0.5°C |
| calibration points: |
0°C/250°C/500°C |
|
Microprocessor precision thermometer Pt100 4-wire, highest accuracy, reference device for any calibrating task (sensor not incl.)
Please order the following accessories separately:
Please order furthermore separately:
= data sheet as *.pdf
= instruction manual as *.pdf
deutsche Beschreibung
Price of the basic article, German V.A.T. incl
:
EUR
247.00
The net price of the basic article GMH 3750
is:
EUR
207.56
|
|
|