GMSD 2,5 MR
for over / under pressure and pressure
difference.
-1,999 bis 2,500 mbar (± 2,500 mbar)
Specification:
| Measuring range: |
-1,999 to 2,500 mbar
(-199,9 ... 250,0 Pa) |
Overload: |
max. 200 mbar |
Resolution: |
0.001 mbar (0.1 Pa) |
Accuracy:
|
(0-2.5mbar) (typ. values)
±0.2%FS (hysteresis and linearity)
+1.0%FS (temperature-influence from 0-50°C) |
Sensor: |
piezo-resistive relative pressure
sensor.
Suitable for air resp. non-corrosive
and non-ionising gases
and liquids. |
|
Pressure connection: |
2 connection pins of nylon for tubes 6 x 1 mm
(6mm outer-diameter and 4 mm inside-diameter. |
Electronics: |
PC board with amplifier
and data memory for sensor data
(meas. range / calibration etc.) integrated in sensor
housing |
Working temperature: |
0 to +50°C |
Relative humidity:
|
0 to +95% r.h. (non-condensing) |
Storage temperature:
|
-40 to +85°C |
Case:
|
ABS plastic with suspension eye
dimensions do not incl. connection pins:
68 x 32.5
x 15 mm (H x W x D)
Dimensions with connection pins: 68 x 32.5 x 27.5mm. |
Unit connection:
|
1 m PVC connection
cable screened
with integral 6-pin mini-DIN-plug, lockable |
Weight:
|
approx. 75g |
GMSD 25 MR
for over / under pressure and pressure difference..
-19.99 to 25.00 mbar (± 25.00 mbar)
Specification:
Meas. range: |
-19.99 to 25.00 mbar
(-1999 ... 2500 Pa) |
Overload: |
max. 300 mbar |
Resolution: |
0.01 mbar (1 Pa) |
Accuracy:
|
(0-25mbar) (typ.
values)
+0.2%FS (hysteresis and linearity)
±0.5%FS (temperature-influence from 0-50°C) |
Other data identical to GMSD 2,5 MR |
GMSD 350 MR
for over / under pressure and pressure difference
-199.9 to 350.0
mbar (± 350.0 mbar)
Specification:
Meas. range:
|
-199.9 to 350.0 mbar
|
Overload: |
max. 1 bar
|
Resolution: |
0.1 mbar |
Accuracy: |
(0-350mbar) (typ. values)
±0.2%FS (hysteresis and linearity)
±0.4%FS (temperature-influence from 0-50°C)
|
Other data identical to GMSD 2,5 MR |
GMSD 7 BA
for absolute pressure
0.00 to 7.00 bar absolute
Specification:
Meas. range: |
0.00 to 7.00 bar absolute |
Overload:
|
max. 10 bar absolute |
Resolution: |
0.01 bar (10 mbar) |
Accuracy: |
typ. values)
(±0.2%FS (hysteresis and linearity)
±0.4%FS (temperature influence from 0-50°C) |
Other data identical to GMSD 1,3 BA |
|
GMSD 2 BR
for over / under pressure and pressure difference
-1,000 to 2,000 mbar
Specification:
Meas. range:
|
-1,000 to 2,000 mbar |
Overload: |
max. 4 bar |
Resolution: |
1 mbar |
Accuracy: |
(0-2,000mbar) (typ.
values)
±0.2%FS (hysteresis and linearity)
±0.4%FS (temperature influence 0-50°C) |
| Other data identical to GMSD 2,5 MR |
GMSD 10 BR
for over / under pressure and pressure difference
-1.00 to 10.00 bar
Specification:
Meas. range: |
-1.00 to 10.00 bar |
Overload: |
max. 10.34 bar |
Resolution: |
0.01 bar (10 mbar) |
Accuracy: |
(0-10bar)
(typ. values)
±0.2%FS (hysteresis and linearity)
±0.4%FS (temperature-influence from 0-50°C) |
Other data identical to GMSD 2,5 MR |
GMSD 1,3 BA
for absolute pressure
0 bis 1300 mbar absolute
Specification:
Meas. range: |
0 to 1,300 mbar absolute
|
Overload: |
max. 4 bar absolute
|
Resolution: |
1 mbar
|
Accuracy: |
(typ. values)
±0.2%FS (hysteresis and linearity)
±0.4%FS (temperature influence from 0-50°C) |
Sensor: |
piezo-resistive absolute pressure
sensor.
Suitable for air resp. non-corrosive
and non-ionising gases and
liquids |
Other data identical to GMSD 2,5 MR |
GMSD 2 BA
for absolute pressure
0 to 2,000 mbar absolute
Specification:
Meas. range: |
0 to 2,000 mbar absolute
|
Overload: |
max. 4 bar absolute
|
Resolution:
|
1 mbar
|
Other data identical to GMSD 1,3 BA |
Options:
|
Sensor for Ex-protection: |
(EEx ib IIC T4 - ZELM 03
ATEX 0136 X) |
Higher probe accuracy: |
by calibration of several
points.
(Additional linearisation points are
memorized in sensor-EEPROM)
-
not possible for GMSD2.5MR and GMSD25MR! |
Certificate of calibration WPD 5 |
(f. IS09000 ff.) incl. several calibration
points of
probe,certificate of calibration:
5 points increase, 5
point decrease |
Certificate of calibration WPD 10 |
(f. IS09000 ff.) incl. several calibration
points of
probe,
certificate of calibration:
10 points increase, 10
point decrease |
|