Pt 100-MEASURING PROBE (conforming to
DIN class B standard
Available for any tolerance or also as Pt1000
upon request and against upcharge
| Ordering type / Range |
Application / Dimensions (mm) / techn. specification |
Response time t90 |
suitable for |
GTF 601
-200 to + 600°C
GTF 601 1/3
-200 to +600°C |
Immersion probe for liquids / gases,
4-wire
handle as per GTF 150, approx. 1m 4-wire PVC cable,
4-pin miniature DIN-type plug, flexible tube, 3 mm average
(smaller tube diameter upon request)
as GTF 601 but 1/3 DIN (+0,1°C at 0°C)
 |
approx. 10 sec. |
GMH 2000 SA
GMH 2000
GMH 35xx
GMH 3750 |
GTF 401
-50 to +400 °C |
Immersion probe for liquids / gases,
4-wire
handle and sensor tube as per GTF 150, approx. 1m 4-wire
PVC cable, 4-pin miniature DIN-type plug

|
approx. 10 sec. |
GMH 2000 SA
GMH 2000
GMH 35xx
GMH 3750
|
|
GES 401
-50 to+ 400°C |
Insertion probe for soft media, 4-wire
handle and sensor tube as per GES 150, approx. 1m 4-wire.
PVC cable, 4-pin miniature DIN-type plug

|
approx. 10 sec. |
GMH 2000 SA
GMH 2000
GMH 35xx
GMH 3750
|
GTF 35
-50 to +400 °C |
Immersion probe for liquids / gases,
4-wire
non-corrosive and acid-resistant stainless steel tube (V4A),
approx. 1 m 4-wire PVC cable,
4-pin miniature DIN-type plug
 |
approx. 10 sec. |
GMH 35xx
GMH 3750
|
|
upcharge for higher accuracy (1/3 DIN +0,1°C
bei 0°C) for Pt 1000 either
(1/10 DIN +0,03°C bei 0°C) (-50 +250°C) for Pt 100
only |
For Pt100 4-wire measuring probe any probe or
cable length possible without re-calibration of the device

ISO 9000 ff manufacturer's certificate of calibration:
Please contact us for other qualifying devices and
probes!
|
Please note:
Pt100 probes are not suitable for mobile and fast surface
measurements. Please make it a rule to use our quick response thermometers
NiCr-Ni (type K):
GTH 1150,
GTH 1170, GMH 1150,
GMH 3200-series with suitable surface
probe.
Immersion probe for liquids and air for GMH3530, Pt100, 4-wire, -50°C to +400°C
= data sheet as *.pdf
deutsche Beschreibung
|