OXY 3610MP - Details

Home --> Devices for Switch Panels / Industry --> Measuring Transducer --> Oxygen Measuring Transducer



 

 

Oxygen measuring transducer for dissolved oxygen in liquids incl. oxygen sensor

OXY 3610 MP

   

Specification:

 
Measuring ranges:

oxygen concentration: 0.00 bis 25.00 mg/l (gelöst)

Temperature: 0.0..50.0°C

 

Accuracy(at nominal temperature 25°C):

Oxygen: +1.5% v. MW + 0,2mg/l

Temperature: +0,1°C + 1 digit

 

Output signal (only O2):

4-20mA (2-wire - standard)

0-10V (3-wire - option)

 

electric isolation:

input electrically isolated

 

auxiliary energy:

12..30 V DC (at output 4-20mA)

18..30 V DC (at output 0-10V - Option)

 

Perm. impedance:

(at 4-20mA): RA[Ohm]= UV[V]-12V /0,02A

Perm. load:

(at 0-10V): RL>3000 Ohm

 

working temperature:

0 to 50°C

storage temperature:

-20..70°C

Relative humidity:

0 to 95% r. H.

 

Reverse voltage protection:

50V permanently

 

Display:

approx. 10mm hohe, 4-digit LCD

 

Housing:

ABS (IP65 - with the exception of sensor plug)

 

Dimensions:

82x80x55mm (without elbow type plug and sensor plug)

 

electric connection:

elbow-type plug acc. to DIN 43650 (IP 65), max. wire coross section: 1.5mm², wire diameter from 4.5 to 7mm

 

Sensor connection:

5pin jack connector, screwable

 

Calibration:

1-point calibration: extremely simple quick calibration in atmospheric air

 

 

oxygen sensor GWO 3600MU:

 
Electrode:

Active membrane type, with integrated NTC-aktiver Membrantyp, with integrated NTC-resistor

 

Response time:

95% in 10sec., depends on the temperature

 

Operation life:

3 years or more, depending on maintenance

 

Operating pressure:

max. 3 bar

 

Flow rate:

min. 30cm/sec.

 

Built in diameter:

d=12,0 + 0,2mm (suitable for 1/2" screw connection)

 

Overall length:

approx. 220 mm (incl. anti-buckling glanding)

 

Built in length:

approx. 110mm

 

Weight:

approx. 180g

 

Connection cable:

approx.  4m, with 5-pin plug, screwable

 

Working temperature:

0 to 40°C

 

Storage temperature:

0 to +60°C

 

 



Oxygen measuring transducer for dissolved oxygen in liquids incl. oxygen sensor, basic price

Please order following accessories separately:



= data sheet as *.pdf

= instruction manual as *.pdf

deutsche Beschreibung

Price of the basic article, German V.A.T. incl : EUR 431.00


The net price of the basic article  OXY 3610MP is:  EUR 362.18 




News